Rigol DM3058 Manuale Utente Pagina 99

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 192
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 98
RIGOL
Programming Guide for DM3058/DM3058E
4-6
Explanations:
The command is valid only when the dB function (CALCulate:FUNCtion DB) is
enabled under calculation is active (CALCulate:STATe ON).
CALCulate:DB:REFerence {<value>|MINimum|MAXimum} - Sets the desired dB
reference value.
CALCulate:DB:REFerence? [MINimum|MAXimum] - Queries the current dB
reference value.
MIN = -120 dB; MAX = 120 dB; DEF = 0 dB.
Return Format:
The query returns the dB reference value.
6. CALCulate:DBM:REFerence
Command Format:
CALCulate:DBM:REFerence {<value>|MINimum|MAXimum}
CALCulate:DBM:REFerence? [MINimum|MAXimum]
Function:
The commands set and query the dBm reference value.
Parameter:
Name Type Range Default
<value>
Consecutive
Integer
2~8000 600
Explanations:
The command is valid only when the dB function (CALCulate:FUNCtion DBM) is
enabled under calculation is active (CALCulate:STATe ON).
CALCulate:DB:REFerence {<value>|MINimum|MAXimum} - Sets the desired
dBm reference value.
CALCulate:DB:REFerence? [MINimum|MAXimum] - Queries the current dBm
reference value.
MIN = 2Ω; MAX = 8000Ω; DEF = 600Ω.
Return Format:
The query returns the current dBm reference value.
Vedere la pagina 98