RIGOL Chapter 2 Command System
Programming Guide for DM3000
2-38
The query returns RISE or FALL.
Default
RISE
8.
:TRIGger:VMComplete:POLAr
Syntax :TRIGger:VMComplete:POLAr?
:TRIGger:VMComplete:POLAr {POSitive|NEGative}
Function Set the polarity of the VMC output at the rear panel.
The query returns POSITIVE or NEGATIVE.
Default
POSitive
9. :TRIGger:VMComplete:PULSewidth
Syntax :TRIGger:VMComplete:PULSewidth?
:TRIGger:VMComplete:PULSewidth
<value>
Function Set the pulse width of the VMC output at the rear panel. The default
unit is ms.
The query returns the pulse width.
Explanation The range of
<value>
is related to the current reading resolution.
DM306x
Reading resolution <value>
4 ½
(ACV/ACI, 3 ½)
1 ms - 29 ms
5 ½ (ACV/ACI, 4 ½) 1 ms - 199 ms
6 ½ (ACV/ACI, 5 ½) 1 ms - 399 ms
DM305x
Reading resolution <value>
3 ¾
(ACV/ACI/DCI, 3 ½)
1 ms - 29 ms
4 ¾
(ACV/ACI/DCI, 4 ½)
1 ms - 199 ms
5 ¾
(ACV/ACI/DCI, 5 ½)
1 ms - 399 ms
Example Set the pluse width to 100 ms:
:TRIGger:VMComplete:PULSewidth 100
The query returns 100.