Rigol DS1000MD Manuale Utente Pagina 70

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 97
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 69
RIGOL
© 2008 RIGOL Technologies, Inc.
Programming Guide for DS1000/DS1000CA Series
2-62
Returned Format:
Query returnsPOSITIVE or NEGATIVE. The double quotes are not returned.
Example:
:TRIGger:VIDEO:POLarityPOSitive Set the Video polarity as POSitive
:TRIGger:VIDEO:POLarity? Query returns POSITIVE
5.3 :TRIGger:VIDEO:STANdard
Command Format:
:TRIGger:VIDEO:STANdard{NTSC|PALSecam}
Explanation:
The command is to set the video type as NTSC or PAL/SECAM.
Quert Format:
:TRIGger:VIDEO:STANdard?
Returned Format:
Query returnsNTSC or PAL/SECAM. The double quotes are not returned.
Example:
:TRIGger:VIDEO:STANdardPALSECAM Set the video type as PAL/SECAM
:TRIGger:VIDEO:STANdard? Query returns PAL/SECAM
5.4 :TRIGger:VIDEO:LINE
Command Format:
:TRIGger:VIDEO:LINE<value>
Explanation:
The command is to set up number of synchronous appointed lines.
The range of <value>: 1~525 for NTSC standard
The range of <value>: 1~ 625 for PAL standard
Query Format:
:TRIGger:VIDEO:LINE?
Returned Format:
Vedere la pagina 69