RIGOL 9 Pass/Fail Test
DS2000 User’s Guide
9-2
To Enable Pass/Fail Test
Press Utility Pass/Fail Enable Test to select ON. Note that when the time
base is in X-Y mode, the Pass/Fail test function can not be enabled.
To start testing, press Enable Test and select “ON”. Then, press Operate to select
to start testing and select to stop testing.
You can select the signal source, set the test mask range, create mask as well as
save and load the test mask. For details, please refer to the following introductions.
To Select Source
Press Source to select the channel (CH1 or CH2) to be tested and only channels
enabled can be selected. During the test, the oscilloscope will judge whether each
frame of waveform in the source complies with the current test mask and those
waveforms pass through the mask area (blue area) is considered as failed.
Mask Range
Users can define their desired test masks.
Press MaskRange MaskRange to select Screen Regionor Cursor Regionfor
test. When Cursor Regionis selected, two gray cursor lines appear on the screen.
At this point, press Cursor A and Cursor B; then, use to adjust the positions of