Service Guide for DM3000 Series
3-5
Table 3-2 Zero Offset of the Standard Equipment (DM306X)
Input
Function
Quick Test
point
Range
Error 1 year
Open
DC Current
2.000000mA
± 1μA
Q
20.00000mA
± 1.2μA
200.0000 mA
± 10μA
1.000000A
± 300μA
10.00000A
[1]
± 2mA
Short
DC Voltage
Q
200.0000mV
± 14μV
2.000000V
± 14μV
20.00000V
± 100μV
200.0000V
± 2.4mV
1000.000V
[2]
± 10mV
Short
2-Wire
Ohms/
4-Wire
Ohms
[3]
200.0000Ω
± 22mΩ
2.000000kΩ
± 40mΩ
20.00000kΩ
± 200mΩ
Q
200.0000kΩ
± 2Ω
1.000000MΩ
± 10Ω
10.00000MΩ
± 600Ω
100.0000MΩ
±15kΩ
Notes:
[1] Continuous current more than DC 7A or AC RMS 7A should be 30 seconds Off and 30 seconds
On;
[2] Each additional volt over ± 500VDC adds 0.02 mV of error;
[3] Errors in above table are corresponding to 2-wire resistance function in math Null. Without
Math Null, the error is ± 0.2 Ω additional errors in 2–wire resistance function;
[4] The reading resolution of multimeter is ;
[5] Q: Optional quick test point.