Chapter 2 Command System RIGOL
DS1000Z Programming Guide 2-105
:TRIGger:RUNT (Option)
Command List:
:TRIGger:RUNT:SOURce
:TRIGger:RUNT:POLarity
:TRIGger:RUNT:WHEN
:TRIGger:RUNT:WUPPer
:TRIGger:RUNT:WLOWer
:TRIGger:RUNT:ALEVel
:TRIGger:RUNT:BLEVel
:TRIGger:RUNT:SOURce
Syntax :TRIGger:RUNT:SOURce <source>
:TRIGger:RUNT:SOURce?
Description
Set or query the trigger source in runt trigger.
Parameter
Name
Type
Range
Default
<source>
Discrete
{CHANnel1|CHANnel2|CHANnel3|CHANnel4}
CHANnel1
Return
Format
The query returns CHAN1, CHAN2, CHAN3 or CHAN4.
Example
:TRIGger:RUNT:SOURce CHANnel2
:TRIGger:RUNT:POLarity
Syntax :TRIGger:RUNT:POLarity <polarity>
:TRIGger:RUNT:POLarity?
Description
Set or query the pulse polarity in runt trigger.
Parameter
Name
Type
Range
Default
<polarity>
Discrete
{POSitive|NEGative}
POSitive
Explanation POSitive: the instrument triggers on the positive runt pulse.
NEGative: the instrument triggers on the negative runt pulse.
Return
Format
The query returns POS or NEG.
Example
:TRIGger:RUNT:POLarity NEGative