RIGOL Chapter 2 Command System
2-132 DS1000Z Programming Guide
:WAVeform Commands
:WAVeform commands are used to read the waveform data and its related settings.
:WAVeform:MODE is
used to set the mode to read the waveform data. In different mode, the parameters definition is different,
as shown in Figure 2-1 and Figure 2-2.
Figure 2-1 Parameters Definition in NORMAL Mode
Figure 2-2 Paramters Definition in RAW Mode
XORigion
YORigion
YREFerence
YINCrement=Verticalscale/25
XINCrement=TimeScale/100
XREFerence
XORigion
YORigion
YREFerence
YINCrement=Vscale/2
XINCrement=1/SampleRate