Chapter 2 Command System RIGOL
Programming Guide for DM3000
2-43
Syntax
:CALCulate:DBM:REFErence?
:CALCulate:DBM:REFErence {
<range>
|MIN|MAX|DEF}
Function
Sets the reference resistance for dBm operations in Ω.
The query returns an integer.
Explanation
<range>
is an integer ranging from 2 to 8000.
The DEF is 600.
15. :CALCulate:LIMIt:STATe
Syntax
:CALCulate:LIMIt:STATe?
:CALCulate:LIMIt:STATe {ON|OFF|1|0}
Function
Turns on or off the Limit operation function.
The query returns ON or OFF.
Default
OFF
16. :CALCulate:LIMIt?
Syntax
:CALCulate:LIMIt?
Function
Queries the current Limit operation result.
The query returns PASS or FAIL.
Explanation
Limit operation function must be turned on before sending this
command.
17. :CALCulate:LIMIt:LOWEr
Syntax
:CALCulate:LIMIt:LOWEr?
:CALCulate:LIMIt:LOWEr {
<range>
|MIN|DEF}
Function
Sets the lower value for Limit operations.
The query returns the lower value of Limit operation in the form of
scientific notation.
Explanation
The lower value should be lower than upper value, for more
details refer to :CALCulate:LIMIt:UPPEr.
Different measurements have different setting ranges:
Measurement
Range
Default
Unit
DC voltage
± 1200
0
V
AC voltage
0 - 900
0
V
DC current
± 12
0
A
AC current
0 - 12
0
A
Resistance
0 - 1.2e+08
0
Ω
Capacitance
0 - 2.4e-04
0
F