RIGOL Chapter 2 Command System
Programming Guide for DM3000
2-44
Frequency
3 - 3.0e+05
3
Hz
Period
3.0e-06 - 3.0e-01
3.0e-06
s
Ratio
± 1.0e+09
-1.0e+09
18. :CALCulate:LIMIt:UPPEr
Syntax
:CALCulate:LIMIt:UPPEr?
:CALCulate:LIMIt:UPPEr {
<range>
|MAX|DEF}
Function
Sets the upper value for Limit operations.
The query returns the upper value of Limit operation in the form of
scientific notation.
Explanation
The upper value should be greater than lower value, for more
details refer to :CALCulate:LIMIt:LOWEr.
<range>
is decided by the current measurement type, for more
details, please refer to the following table.
Measurement
Range
Default
Unit
DC voltage
± 1200
1
V
AC voltage
0 - 900
1
V
DC current
± 12
1
A
AC current
0 - 12
1
A
Resistance
0 - 1.2e+08
1
Ω
Capacitance
0 - 2.4e-04
1
F
Frequency
3 - 3.0e+05
3.0e+05
Hz
Period
3.0e-06 - 3.0e-01
3.0e-01
s
Ratio
± 1.0e+09
0
Vedere la pagina 55