Example:
:TRIGGER:ALTERNATION:VIDEO:LINE100, SOURceB
Set the specified line number as 100.
:TRIGger:ALTernation:VIDEO:LINE?SOURceB Return 100.
(15) :TRIGger:ALTernation:COUPling
Command Format:
:TRIGger:ALTernation:COUPling{DC | AC | HF | LF}[,<source>]
Function:
The command sets the coupling mode.
DC (Allow all signals pass);
AC (Reject DC signals and attenuate AC signals below 10Hz).
HF (Reject high frequency signals above 150 kHz).
LF (Reject DC and attenuate low frequency signals below 8kHz).
The <source> may be SOURceA or SOURceB, and the source A and B are
varying with the current alternation channel.
Query Format:
:TRIGger:ALTernation:COUPling?[<source>]
Return Format:
The query returns DC, AC, HF or LF.
Example:
:TRIGger:ALTernation:COUPlingDC, SOURceB
Set the couping mode as DC.
:TRIGger:ALTernation:COUPling?SOURceB Return DC.
(16) :TRIGger:ALTernation:SENSitivity
Command Format:
:TRIGger:ALTernation:SENSitivity<count>[,<source>]
Function:
The command sets the trigger sensitivity of alternation trigger, the count
range is 0.1div~1div. The <source> may be SOURceA or SOURceB, and
the source A and B are varying with to the current alternation channel.
Vedere la pagina 81