2 Command System RIGOL
MSO2000A/DS2000A Programming Guide
2-337
:TRIGger:VIDeo:LINE
Syntax
:TRIGger:VIDeo:LINE <line>
:TRIGger:VIDeo:LINE?
Description
Set or query the line number in video trigger when the sync type is Line (refer to
:TRIGger:VIDeo:MODE
command).
Parameter
Name Type Range Default
<line> Integer
NTSC: 1 to 525
PAL: 1 to 625
480P: 1 to 525
576P: 1 to 625
720P60HZ: 1 to 750
720P50HZ: 1 to 750
720P30HZ: 1 to 750
720P25HZ: 1 to 750
720P24HZ: 1 to 750
1080P60HZ: 1 to 1125
1080P50HZ: 1 to 1125
1080P30HZ: 1 to 1125
1080P25HZ: 1 to 1125
1080P24HZ: 1 to 1125
1080I30HZ: 1 to 1125
1080I25HZ: 1 to 1125
1080I24HZ: 1 to 1125
1
Return Format
The query returns the line number in video trigger when the sync type is Line in integer.