Programming Guide for DM3000 Series
Function:
The commad queries two circuits resistance.
Return Format:
The query returns the current resistance in the form of 8.366031e-05, and unit is Ω.
(for example)
32. :measure:resistance
Syntax:
:measure:resistance□{<
range
>|MIN|MAX|DEF}
Function:
The command sets the measure range of two circuits resistance.
Parameters :
Explanations:
The table 2-6 indicates the parameters and the range.
Table 2- 6 The two circuits resistance range
Example:
To set the measure range of the two circuits resistance as 200.0Ω:
:measure:resistance□MIN or :measure:resistance□0