40. :measure:frequency
Syntax:
:measure:frequency{<
range
>|MIN|MAX|DEF}
Function:
The command sets input voltage range in measuring frequency.
Parameters:
Name
Range
Default
<
range
>
{0|1|2|3|4}
2
Explanations:
The frequency range is 3Hz300kHz.The Table 2- 3 indicates the parameters and the
ranges.
Example:
To set the voltage range as 20V:
:measure:frequencyDEF
41. :measure:frequency:range?
Syntax:
:measure:frequency:range?
Function:
The command queries AC voltage range in measuring frequency.
Explanations:
The range is the same as the DC voltage. See Table 2- 3.
Return Format:
The query returns the current range parameter: 0123 or 4.( see Table 2- 3)