Rigol DM3051 Specifiche Pagina 38

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 123
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 37
RIGOL
© 2008 RIGOL Technologies, Inc.
Programming Guide for DM3000 Series
2-26
Function:
The command queries four circuits resistance.
Return Format:
The query returns the current four circuits resistsnce in the form of 8.822946e-05, and
unit is Ω. (for eample)
36. :measure:fresistance
Syntax:
:measure:fresistance{<
range
>|MIN|MAX|DEF}
Function:
The command sets the measure range of four circuits resistance.
Parameters:
Name
Range
Default
<
range
>
{0|1|2|3|4|5|6 }
3
Explanation:
The Table 2- 6 indicates the parameters and their meanings.
Example:
To set the measure range of the four circuits resistance as 200.0Ω:
:measure:fresistanceMIN or :measure:fresistance0
37. :measure:fresistance:range
Syntax:
:measure:fresistance:range?
Function:
The command queries the measure range for four circuits resistance.
Return Format:
The query returns the current range parameter: 0 12 3 4 5 or 6.(see
Vedere la pagina 37