RIGOL
Programming Guide for DM3058
Return Format:
The query returns max transient voltage in DCV measurement.
8. FETCh:VOLTage[:DC]:PEAK:MINimum?
Command Format:
FETCh:VOLTage[:DC]:PEAK:MINimum?
Function:
The command queries min transient voltage in DCV measurement.
Explanations:
Before sending the command, please enable DC voltage measuring function, and
enble DCV peak-peak measuring function with the command [SENSe:]
VOLTage[:DC]:PEAK:STATeON.
Return Format:
The query returns min transient voltage in DCV measurement.
9. FETCh:VOLTage[:DC]:PTPeak?
Command Format:
FETCh:VOLTage[:DC]:PTPeak?
Function:
The command queries the difference between the highest and lowest DC
transient voltage.
Explanations:
Before sending the command, please enable DC voltage measuring function, and
enable DCV peak-to-peak measuring function with the command [SENSe:]
VOLTage[:DC]:PEAK:STATeON.
Return Format:
The query returns the peak-to-peak value in DCV measurement.