Programming Guide for DM3058
5-53
Return Format:
The query returns the enable state of DCI interval time function.
9. [SENSe:]CURRent[:DC]:NPLC
Command Format:
[SENSe:]CURRent[:DC]:NPLC{<PLCs>|MIN|MAX|DEF}
[SENSe:]CURRent[:DC]:NPLC?[{MIN|MAX}]
Function:
The command sets or queries the interval time of DCI measurement, input values
and returned values are both multiples of PLC.
Parameters:
Name
Type
Range of Values
Default
<PLCs>
Discrete
{0.006|0.02|0.06|0.2|1|2|10|100}
1 PLC
Explanations:
[SENSe:]CURRent[:DC]:NPLC{<PLCs>|MIN|MAX|DEF}: Set the interval time
of DCI measurement, input values is multiples of PLC. <PLCs> may be 0.006,
0.02, 0.06, 0.2, 1, 2, 10 or 100; MIN = 0.006 PLC; MAX = 100 PLC.
[SENSe:]CURRent[:DC]:NPLC?[{MIN|MAX}]: Query the interval time of DCI
measurement, returned values is multiples of PLC.
Return Format:
The query returns the interval time of DCI measurement which is multiples of
PLC.
10. [SENSe:]CURRent[:DC]:NULL[:STATe]
Command Format:
[SENSe:]CURRent[:DC]:NULL[:STATe]{ON|OFF}
[SENSe:]CURRent[:DC]:NULL[:STATe]?
Function:
The command sets or queries the enable state of DCI NULL function.
Vedere la pagina 168