Programming Guide for DM3058
[SENSe:]RESistance:ZERO:AUTO{ON|OFF}: Enable or disable 2WR autozero
measuring function.
[SENSe:]RESistance:ZERO:AUTO?: Query the enable state of 2WR autozero
measuring function.
Return Format:
The query returns the enable state of 2WR autozero measuring function.
43. [SENSe:]FRESistance:APERture
Command Format:
[SENSe:]FRESistance:APERture{<seconds>|MIN|MAX|DEF}
[SENSe:]FRESistance:APERture?[{MIN|MAX}]
Function:
The command sets or queries the interval time of 4WR measurement.
Parameters:
Name
Type
Range of Values
Default
<seconds>
Consecutive
Real Number
400 ms~2 s
400 ms
Explanations:
[SENSe:]FRESistance:APERture{<seconds>|MIN|MAX|DEF}: Set the interval
time of 4WR measurement. Unit is s, so 400 ms should be input in form of 0.4.
[SENSe:]FRESistance:APERture?[{MIN|MAX}]: Query the interval time of 4WR
measurement.
2-wire and 4-wire resistance function both are influenced by the command. All
the command of 2WR and 4WR will have influence over the setting of resistance.
Return Format:
The query returns the interval time of 4WR measurement.
44. [SENSe:]FRESistance:APERture:ENABled
Command Format:
Vedere la pagina 189