5-101
Parameters:
Name
Type
Range of Values
Default
<seconds>
Consecutive
Integer
0~3600
Required
parameter
Explanations:
TRIGger:DELay{<seconds>|MIN|MAX|DEF}: Set the delay time.
TRIGger:DELay?[{MIN|MAX}]: Queries the delay time between trigger signal
and starting measurement.
In continuity test and diode test, delay time is not set.
For AC measurement, the delay time is affected by bandwidth.
The command ―CONFigure‖ or ―MEASure?‖ can be used to set automatically delay
time.
The command ―TRIGger:DELay‖ can be used to set delay time as well as disable
auto delay function (TRIGger:DELay:AUTO).
It is suggested that the command ―SENSe:<function>:RANGe:AUTOOFF‖
should be used to disable autorange function, the command ―SENSe:<function>:
RANGe‖, ―CONFigure‖ or ―MEASure?‖ be uesd to set range by manual.
Return Format:
The query returns the delay time between trigger signal and starting
measurement.
3. TRIGger:DELay:AUTO
Command Format:
TRIGger:DELay:AUTO{ON|OFF}
TRIGger:DELay:AUTO?
Function:
The command sets or queries the enable state of trigger autodelay function.
Explanations:
TRIGger:DELay:AUTO{ON|OFF}: Enable or disable trigger autodelay function.
TRIGger:DELay:AUTO?: Query the enable state of trigger autodelay function.
The command TRIGger:DELay can be used to set delay time as well as disable
auto delay function (TRIGger:DELay:AUTO).