Programming Guide for DM3058
2-4
4. *ESR
Command Format:
*ESR
Function:
The command queries the event value of Event Status Register (ESR).
Return Format:
The query returns the event value of ESR.
5. *PSC
Command Format:
*PSC{0|1}
*PSC?
Function:
The command sets or queries the value of ―power-on status clear (PSC)‖.
Explanations:
*PSC{0|1}: Enable or disable the clear operation of register as soon as the
instrument is power-on. ―*PSC 0‖ indicates disable, that is the enable value of
every register keeps the value before power-off; ―*PSC 1‖ indicates enabler, that
is all the enable values are reset.
*PSC?: Query the value of ―power-on status clear‖.
Return Format:
The querry returns 0 or 1. ―0‖ denotes that the register will not been cleared
when power-on; ―1‖ denotes the register will be cleared when power-on.