Rigol DM3058 Specifiche Pagina 181

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 264
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 180
RIGOL
Programming Guide for DM3058
5-65
Explanations:
[SENSe:]VOLTage[:DC]:PEAK:STATe {ON|OFF}: Enable or disable DCV
peak-to-peak measuring function.
[SENSe:]VOLTage[:DC]:PEAK:STATe?: Query the enable state of DCV
peak-to-peak measuring function.
The command [SENSe:]VOLTage[:DC]:NULL:VALue can be uesd to set DCV
peak-to-peak value.
Return Format:
The query returns the enable state of DCV peak-to-peak measuring function.
29. [SENSe:]VOLTage[:DC]:RANGe:AUTO
Command Format:
[SENSe:]VOLTage[:DC]:RANGe:AUTO{ON|OFF}
[SENSe:]VOLTage[:DC]:RANGe:AUTO?
Function:
The command sets or queries the enable state of DCV autorange measuring
function.
Explanations:
[SENSe:]VOLTage[:DC]:RANGe:AUTO {ON|OFF}: Enable or disable DCV
autorange measuring function. If DCV peak-to-peak measuring function is ON,
the measuring range will be selected according to input signal.
[SENSe:]VOLTage[:DC]:RANGe:AUTO?: Query the enable state of DCV autorange
measuring function.
Return Format:
The query returns the enable state of DCV autorange measuring function.
30. [SENSe:]VOLTage[:DC]:RANGe[:UPPer]
Command Format:
[SENSe:]VOLTage[:DC]:RANGe[:UPPer]{<range>|MIN|MAX|DEF}
[SENSe:]VOLTage[:DC]:RANGe[:UPPer]?[{MIN|MAX}]
Vedere la pagina 180
1 2 ... 176 177 178 179 180 181 182 183 184 185 186 ... 263 264

Commenti su questo manuale

Nessun commento